T10000 - 1.46.x09 Diagnostic Bug - Can't Run Offline Diags Load-Unload Test While Drive is in a Library
Related Items |
- Sun StorageTek T10000A Tape Drive
- Sun StorageTek T10000B Tape Drive
|
Related Categories |
- PLA-Support>Sun Systems>TAPE>Tape Hardware>SN-TP: STK T-Series Drive
|
T10K 1.46.x09 Diagnostic Bug
In this Document
Symptoms
Changes
Cause
Solution
Symptoms
When
attempting to run the load/unload test it fails as follows:
8:10 PM Aug 17, 2010 - START C:\Program Files\VOP
1.0.15\diags\Load_Unload_Test.117 Test
Install a DUMP tape for Load_Unload_Test
8:11 PM Aug 17, 2010 - Tape Cartridge is LOADED
DC_BUFFER_TEST PASSED
DUMP tape was successfully loaded
UNLOADING
8:12 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:12 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOADING
EPT_DIAG_FAILED_LOAD
I then loaded 1.44.210 on the drive and re-ran the test:
8:19 PM Aug 17, 2010 - START C:\Program Files\VOP
1.0.15\diags\Load_Unload_Test.117 Test
Install a DUMP tape for Load_Unload_Test
8:19 PM Aug 17, 2010 - Tape Cartridge is LOADED
DC_BUFFER_TEST PASSED
DUMP tape was successfully loaded
UNLOADING
8:20 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:20 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOADING
8:20 PM Aug 17, 2010 - Tape Cartridge is LOADED
UNLOADING
8:20 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:20 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOADING
8:21 PM Aug 17, 2010 - Tape Cartridge is LOADED
UNLOADING
8:21 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:21 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOADING
8:21 PM Aug 17, 2010 - Tape Cartridge is LOADED
UNLOADING
8:22 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:22 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOADING
8:22 PM Aug 17, 2010 - Tape Cartridge is LOADED
UNLOADING
8:22 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:22 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOAD_UNLOAD_TEST ran for 3 minutes and 15 seconds
FDE_DONE: Load_Unload_Test.tcf test has completed
8:22 PM Aug 17, 2010 - END C:\Program Files\VOP
1.0.15\diags\Load_Unload_Test.117 Test
I then reloaded 1.46.209 on the drive and ran the test again:
8:27 PM Aug 17, 2010 - START C:\Program Files\VOP
1.0.15\diags\Load_Unload_Test.117 Test
Install a DUMP tape for Load_Unload_Test
8:27 PM Aug 17, 2010 - Tape Cartridge is LOADED
DC_BUFFER_TEST PASSED
DUMP tape was successfully loaded
UNLOADING
8:28 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:28 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOADING
EPT_DIAG_FAILED_LOAD
The drive just sits there thinking it's running the test. (If I press
"Done" it says it's still running diags) If I interrupt the test by
removing the tape from the elevator and reinserting it I get the following:
8:30 PM Aug 17, 2010 - Tape Cartridge is EJECTED
LOAD failed with FSC: #000032B8
8:30 PM Aug 17, 2010 - Tape Cartridge is LOADED
UNLOADING
8:30 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
8:30 PM Aug 17, 2010 - Tape Cartridge is UNLOADED
LOAD_UNLOAD_TEST ran for 3 minutes and 24 seconds
FDE_DONE: Load_Unload_Test.tcf test has completed
Changes
1.46.109 (T10000A) or 1.46.209 (T10000B) code was installed.
Cause
Starting with 1.46.109/209, you can't run FDE Offline Diags Load-Unload
test on a
drive that's in a library tray AND has the TTI cable (P40) connected to
the drive, EVEN
if the drive is on a bench.
Solution
This means that for SL8500 drives in a tray AND on a bench, you'll have
to remove the
top lid and disconnect the TTI cable/P40 (for L700 and SL3000 just pull
the TTI/P40
on the back of the drive).
Apparently there was an issue with t10k drives and 1.44 110/210 where
sometimes drives
were not being 'seen' by the library after drive ipl, so the 'user' had
to force an 'error' on
the drive so that the library would 'see' the drive. That fix now means
you have to unplug
TTi/P40 to run the FDE Offline Diags Load-Unload test on trayed drives
running 1.46.109/209. Apr 12 2012 update: Engineering has agreed to change the code to allow running of this diagnostic without having to unplug the TTI cable from the drive. Reference CR 7160658.