Solution Type Technical Instruction Sure Solution 1014191.1 : C4 - How to Calibrate Camera in Library
PreviouslyPublishedAs 220552 Oracle Confidential (PARTNER). Do not distribute to customers Reason: Confidential information provided to Oracle and Partner Service Personnel
Applies to:Sun Storage C4 Tape LibrarySun Storage C4 Tape Library - Version: Not Applicable and later [Release: N/A and later] All Platforms Checked for relevance on 4-Jan-2011. GoalHow to Calibrate Camera in C4 Library.SolutionSteps to Follow:Any time a hand/camera assembly is to be changed out for target calibration errors or when a new hand/camera assembly is placed into the C4 library, it is highly encouraged to run the Camera Calibration Routine to make sure the camera is lined up properly. Download the attached Camera Calibration tool and instructions. The camera is NOT a FRU. If the camera fails calibration, the whole robotic hand assembly must be replaced. How to calibrate camera in library, How to avoid unnecessary camera replacement, PX500, Camera, Camera calibration procedure, Calibration tool, Internal target failure, Mode sense error, Critical event code C2502410, Initialization failure, C4, 4/44/00, Hand replacement Previously Published As STKKB80136 Attachments This solution has no attachment |
||||||||||||
|